Laboratories

GPPS-evo v4 Laboratory Gas Pressure Proving System

The GPPS-evo v4 is a laboratory gas interlocking system with the ability to monitor up to sixteen gas sensors (w/power pack).

The Gpps-evo v4 is a gas pressure proving and gas detection system designed for use primarily in education laboratories.

Gas Pressure Proving

Using our patented differential pressure proving method the performs a downstream integrity check of pipework and outlets which could be left open to atmosphere. Able to test a variety of gas delivery lines including oxygen up to 10bar It also monitors for low pressure and overpressure without opening the solenoid valve, making it the only truly safe method of proving available.

see Why choose Medem differential pressure proving?

Gas Detection

Up to 16 gas detectors can be connected to the system, low level alerts will cause a audible tone and visual message on screen, high level alerts will additionally isolate the gas supply.

In-line with the updated legislation of BB101 and IGEM/UP11/Edition 3 the system can also monitor the CO2 level when fitted with a Medem detector displaying the PPM.

CO2 is considered a primary indicator of air quality, with the latest legislation setting maximum limits for levels within educational facilities.

Carbon MonoxideNatural Gas and Oxygen detectors can also be monitored.

An auto-stop timer is incorporated to help prevent unauthorised use of the gas out-of-hours.

There are several ‘engineer functions’ which can be used to aid in the set-up of the system.

Remote Indications

All system states and alerts can be remotely monitored and indicated via our “BMS” relays. There are five potential free relays each of which can configured separately.

  • Gas Pressure Proving
  • Gas detection (1-16 detectors)
  • Carbon Dioxide Level Monitoring
  • Fan interlocking
  • Event log
  • Fire test isolation mode
  • Two settable SELV relays
  • Low and high incoming gas pressure monitoring alerts
  • Engineer functions

The panel can monitor up to 6 detectors directly and up to 16 when installed with an AD-PP power pack. The current range is:

Click here to find out more about our product warranty

A guide to our method of gas pressure proving.
Power230 VAC – 3 amp fused spur.
Product Dimensions255mm x 180mm x 80mm
Weight1.4kg
Temperature Range-5° to 55°C.
Humidity20% to 90%RH Non-Condensing.
Ingress ProtectionIP64
Housing MaterialABS Plastic
ColourGrey
MountingWall Mount
SensorsPellistor, Toxic, Oxygen, NDIR.
Outputs & Connections1x 16 channel VNOM power & sensor data .
2x Potential free change over relays. (Mains rated)
3x Potential free change over relays. (SLEV)
1x 12volt supply rails.
1x Sender unit (2core low voltage).
1x 0-10v fan speed.
1x Remote emergency stop N/C.
1x Fire Alarm N/C.
2x Fan PD switches.

Alarm Types

Fixed low and high alarm peak levels, high alarm activation will latch and isolate the connected gas valve.

Peak Alarms

Methane/Natural Gas (CH4) & Propane (C3H8)

  • 5% Low alarm
  • 10% High alarm

Carbon Monoxide (CO)

  • 20PPM Low Alarm
  • 30PPM High Alarm

Carbon Dioxide (CO2) – Selectable

  • 1500PPM
  • 2200PPM
  • 2800PPM
  • 4000PPM
  • 5000PPM

Oxygen (O2) Depletion

  • 19.5% Pre Alarm
  • 18% Full Alarm

All our products come with a 5-year warranty

Medem’s industry leading five-year warranty is a demonstration of the company’s confidence in the quality and design of its products.

This applies to the main panels and sender units, gas detectors carry a 2-year warranty.

Why not let us commission your system and receive a 10-year Warranty?

Where we’ve commissioned an installed system, the warranty is extended from five to ten years. At the time of commissioning, photos and gas pressure readings are recorded along with other relevant details of the installation. A 24-hour helpline number and the serial number is attached with the panel and all details are then entered on to the Medem database.

Service

The panel and sensors must be re-commissioned and tested every 12 months as a minimum but with consideration of every 6 months dependent on gas type and environment. Medem offers cost-effective yearly service plan options which include detector exchange

Related Products

Carbon Monoxide (CO) Gas Detector

Carbon Dioxide (CO2)

  • Gas detector
  • Designed for use with Medem systems

View Product
SEC-L v5 Multi-function Laboratory Services and Environmental Control System

SEC-L v5 Gas Proving and Interlock System for Laboratory Services

  • Designed for Laboratory
  • Gas pressure proving system
  • Occupancy based ventilation control using CO₂ detectors
  • Electrical and water isolation

View Product
SEC-Elite Programmable multi-line Gas Proving Detection System

SEC-Elite Multi-line Gas Proving Detection System

  • Designed for Laboratories
  • Multi Gas Proving system for up to 4 different gas lines
  • Gas Detection
  • Electrical and water isolation
  • Fully programmable system

View Product
CM2L-K Two Channel Current Monitor with LCD Display

CM2L-K Two Channel Current Monitor with LCD Display

  • Designed for Laboratories
  • Monitor two fans
  • Fan current load display on screen

View Product